UCC T5 installation

Dimensions

UCC dimensions - generic

NOTE: Dimensions in mm (in).


Weight

2.1 kg (4 lb 10 oz)


UCC T5 can either be free-standing or used in a 19 inch rack system.

WARNING: Ensure the UCC T5 is disconnected from the mains supply during installation.

Free-standing installation

The UCC T5 unit draws air from the right hand side when viewed from the front and expels air out of the left hand side. A minimum clearance gap of 10 mm is necessary between the sides of the unit and any potential obstruction.

Controller dimensions and air direction diagram

Mounting in a 19 inch rack

The rack mounting kit (part number A-5518-0005) contains two brackets and four M5 × 6 mm screws.

Assemble the brackets to the UCC T5 as shown below:

CMM controller mounting

Cable lengths

UCC T5 to SPA3-2 connection

The units must be linked by the supplied shielded CAT 5E 300 mm cable(s), no other cables are to be used.

Ethernet cable link to PC

This is a standard ethernet CAT 5E cable, a 5 m cable is supplied as part of the CMM controller kit. Lengths up to 100 metres can be used.