ACR3 troubleshooting
Black deposit on probe autojoint
Possible cause | Solution |
---|---|
Badly aligned probe head. | Re-align probe head - refer to 'Head alignment for ACR3'. |
Badly aligned ACR3. | Re-datum ACR3 - refer to 'Datuming the ACR3'. |
Excessive wear on ACR3 port
Possible cause | Solution |
---|---|
Badly aligned probe head. | Re-align probe head - refer to 'Head alignment for ACR3'. |
Badly aligned ACR3. | Re-datum ACR3 - refer to 'Datuming the ACR3'. |
Approaching the expected life of the ACR3 port. | Replace the port - refer to 'ACR3 accessories and spare parts' and 'Maintenance'. |
ACR3 does not run smoothly
Possible cause | Solution |
---|---|
Badly aligned probe head. | Re-align probe head - refer to 'Head alignment for ACR3'. |
Badly aligned ACR3. | Re-datum ACR3 - refer to 'Datuming the ACR3'. |
Damage occured to ACR3. | Return the ACR3 to the local Renishaw service centre. |
Head obstructing during ACR3 lock / unlock sequence
Possible cause | Solution |
---|---|
Badly aligned probe head. | Re-align probe head - refer to 'Head alignment for ACR3'. |
Badly aligned ACR3. | Re-datum ACR3 - refer to 'Datuming the ACR3'. |
Attempting to "stack" extension bars. | The ACR3 system has not been designed to support the stacking of extension bars. |
Damage occured to ACR3. | Return the ACR3 to the local Renishaw service centre. |