Skip navigation

SCP600 pick up / put down routines

NOTE: The following section assumes that the datum created in 'Datuming the SCP600' is the axis system that the machine is working in.

Put down routine

The following table lists the recommended procedure for putting down a stylus module on the SCP600. All dimensions are absolute and given in mm.


Operation

X-axis

Y-axis

Z-axis

Inhibit the probe and move to stand-off position

0

50

8

Move into port

0

0

8

Detach stylus

0

0

0

Move to stand-off position

0

50

0


Pick up routine

The following table lists the recommended procedure for picking up a stylus module from the SCP600. All dimensions are absolute and given in mm.


Operation

X-axis

Y-axis

Z-axis

Move to stand-off position

0

50

8

Move into port

0

0

8

Attach stylus

0

0

0

Move to stand-off position and activate the probe

0

50

0