MH20 and MH20i fault finding
MH20 and MH20i
Poor measuring performance:
Possible cause | Remedy |
---|---|
Loose mounting. | Ensure shank mounting screws are tight and mounting to CMM is secure. |
Problem with TP20 probe module (see 'TP20 installation'). | Diagnosis of the probe unit should be carried out with axes correctly locked and without head adjustment between probe points. |
Problem with swivelling / indexing unit. | Diagnosis of indexing unit should only be carried out following confirmation of satisfactory probe performance. |
No probe signal and / or no probe status LED:
Possible cause | Remedy |
---|---|
Cable faulty / not connected. | Check continuity of cabling from head to interface / machine control. |
Probe interface faulty / not connected. | Ensure correct connection of interface/machine control. |
Poor repeatability (MH20i only):
Possible cause | Remedy |
---|---|
Axes lock procedure incorrect. | Ensure lock lever is fully rotated to lock position. |
Axes unlock procedure incorrect. | Ensure lock lever is fully rotated to unlock position during indexing. |
Forces imparted onto head during lock-up. | Unlock and re-lock holding lock lever only. |
Attempted lock-up in incorrect un-qualified position. | Unlock, reposition correctly and re-lock. |
Axes ‘rattling' during indexing (MH20i only):
Possible cause | Remedy |
---|---|
Incorrect unlock procedure. | Ensure lock lever is fully rotated to unlock position. |
Incorrect indexing procedure. | Index each axis separately. |
TP20 module
Poor measuring performance:
Possible cause | Remedy |
---|---|
Stylus configuration too long or not rigid. | Use shorter stiffer stylus configuration. |
Poor stylus assembly. | Ensure stylus joints are kept to a minimum and that all joints are clean and secure. |
Contamination / damage of stylus ball. | Inspect for damage, clean thoroughly with solvent. |
Trigger force too high. | Use lower force module. |
Unwanted triggering during probe or CMM movement:
Possible cause | Remedy |
---|---|
Trigger force too low / stylus configuration too heavy. | Use higher force module / reduce mass of stylus configuration. |
Probe fails to rearm after trigger:
Possible cause | Remedy |
---|---|
Trigger force too low / stylus configuration too heavy. | Use higher force module / reduce mass of stylus configuration. |
Probe reseat failure. | Re-trigger probe. If problem persists please return to Renishaw for service. |
Loss of measuring accuracy:
Possible cause | Remedy |
---|---|
Mounting not secure. | Check that the MH20i is correctly mounted on the shank and that the screws are secure. Check clamping mechanism in CMM quill is secure for MH20 and MH20i. |
MH20 / MH20i not fully locked. | Ensure that the thumbwheel / lock lever is turned fully clockwise. |
Force imparted to probe module after locking. | Re-qualify the probe. |
MH20 / MH20i worn or damaged. | Use only with the specified probe and extension combinations. |
Faulty probe module. | Check by substitution or return to Renishaw or your supplier. |