CMM controller cabinet maintenance

Authorised personnel service and maintenance

NOTE: There is hazardous live 240 Vac present within the cabinet.

CAUTION: Ensure that mains is fully isolated before any service or maintenance operations are carried out.

CAUTION: Following any service and maintenance an inspection of the mains wiring and protective bonding should be carried followed by a PAT test.

Periodic checks

It is advised to periodically check the cabinet and cable management system for damage or wear to key parts. Any visible damage or wear should be reported to the system maintenance engineer.

User serviceable parts

The cabinet has two fans which are the only user serviceable components.

Cooling fans

CAUTION: Ensure the cabinet is switched OFF before changing the filter.

The two cabinet fans are used to push air out of the cabinet for cooling the equipment within. One is located on each lower side panel.

Cabinet - cooling fans

The fans have mesh filters. Renishaw recommends that these are changed at least every 12 months. If the kiosk is located in an environment with high levels of particulates in the air, the filters should be changed at more regular intervals.

Description

Part number

Mesh filter replacement kit (pack of five)

A-6509-0400

These filters should only be ordered from Renishaw.

Cleaning

The CMM controller cabinet should be cleaned with a dry, lint-free cloth.