Skip navigation

PH10M PLUS / PH10MQ PLUS motorised indexing probe head

Technical specifications
Documents
Part numbers
Additional information

PH10M PLUS shank-mounted and PH10MQ quill-mounted heads with multiwired probe capability and a Renishaw autojoint

  • PH10M PLUS and PH10MQ PLUSThe PH10M PLUS and PH10MQ PLUS are motorised indexing heads that re-orientate the probe repeatably to any one of 720 positions at 7.5° increments
  • Autojoint mount for repeatable probe positioning
  • Torque capable of lifting a 300 mm (11.81 in) long extension bar
  • Full multiwire capability for probes such as TP7M and SP25M
  • Rapid automatic probe interchange via ACR1 or ACR3
  • AM1 for PH10M PLUS and AM2 for PH10MQ PLUS adjustment modules for aligning the probe head to machine
  • PH10MQ PLUS fits inside the quill for a neater and shorter probe mount

Dimensions

PH10M PLUS dimensions


PH10MQ PLUS dimensions

NOTE: Dimensions in mm (in).

PH10M PLUS - A and B-axisPH10MQ PLUS - A and B-axis

Specification


Repeatability of position

<0.4 μm (0.00002 in) @ 100 mm from centre of A-axis rotation

Positioning

Step

7.5°

A-axis 0° to 105°

15 positions

B-axis -180° to +180°

48 positions

Total

720 positions

Probe mounting

PH10T PLUS

M8 thread, 18 mm diameter bush face

PH10M PLUS / PH10MQ PLUS

Multiwired autojoint

Dimensions from quill face

XY (all)

62 mm

Z (PH10T PLUS)

102 mm

Z (PH10M PLUS)

117 mm

Z (PH10MQ PLUS)

73 mm

Weight

PH10T PLUS

595 g

PH10M PLUS

620 g

PH10MQ PLUS

730 g

System operating temperature

+10 °C to +40 °C (+50 °F to +104 °F)

System storage temperature

-10 °C to +70 °C (+14 °F to +158 °F)


The PHC10-3 PLUS is powered from the ac mains supply via an IEC 320 connector.

The operating voltages of the unit are as follows:

100 Vac - 240 Vac +10%, -15% 47 Hz - 66 Hz, 30 W maximum

Environmental conditions

The following environmental conditions comply with those defined in BS EN61010 - 1:2001:


Indoor use

IP30 (no protection against liquids)

Altitude

Up to 2000 m

Operating temperature range

0 °C to +50 °C

Storage temperature range

-10 °C to +70 °C

Relative humidity

80% maximum for temperatures up to +31 °C

Pollution degree

2