Navigation
Module setting tool (MST)
Using the module setting tool
1. Place the SFH holder within the MST clips so that the knuckle adjustment screw is accessable via the hole in the MST.
2. Pass the 2 mm hexagonal key through the hole and loosen the knuckle joint enough to freely move the SFM module.
3. Position the SFM module to the required angle using the alignment mark as a guide.
4. Tighten the knuckle joint using a torque of 0.4 Nm to 0.5 Nm, then remove the MST from SFH holder.