Navigation
RCP TC and RCP TC-2 LED status indication and troubleshooting
RCP TC
LED | Status | Troubleshotting |
---|---|---|
Green | Power on, probe present | n/a |
Amber | Power on, no probe | If the probe is present, clean the port and probe contacts. |
Red | Error | Cycle power to restart. If the problem persists contact your supplier. |
RCP TC-2
LED | Status | Troubleshooting |
---|---|---|
Green | Power on, probe present, temperature within target | n/a |
Flashing green | Power on, probe present, temperature outside of target | Wait for the rack to reach stable temperature. |
Amber | Power on, probe not present, temperature within target | If the probe is present, clean the port and probe contacts. |
Flashing amber | Power on, probe not present, temperature outside of target | If the probe is present, clean the port and probe contacts. Wait for the rack to reach stable temperature. |
Flashing red | Error | Cycle power to restart. If the problem persists, contact your supplier. |
WARNING: The RCP TC or RCP TC-2 plate may be hot.