Navigation
PH10 Trigger Force Gauge (gram gauge)
Probe trigger force is critical to ensure accuracy and reliable operation, this can be easily measured and set using the gram gauge.
It is strongly recommended that the gram gauge is used when adjustments are to be made to probe trigger force.
For further information on the trigger force gauge, please refer to the trigger force gauge guide (Renishaw part number H-1000-4054).
NOTE: To convert mass in grams to trigger force in Newtons the following equation is used:
Trigger force in Newtons = Mass in grams/100
For example: 8 grams = 0.08 N /100